Theory, Practice, Technical Preconditions, Advantages, Foreign
Gas
Accumulation
Jan A. Baum
Department of Anaesthesia and Intensive Care
Hospital St. Elisabeth-Stift
1. Short history of the rebreathing technique in anaesthesiology
As early as in 1850 John Snow recognized that a considerable amount of inhalation anaesthetics were exhaled unchanged in the expired air of anaesthetized patients. He concluded and could proof that the narcotic effect really could be markedly prolonged by reinhaling these unused vapours (1). About 75 years later, in 1924, rebreathing systems equipped with carbon dioxide absorbers were introduced into anaesthetic practice. While Ralph Waters used a to-and-fro system (2), the German gynecologist Carl J. Gauss and the chemist Hermann D. Wieland did advocate the use of a circle system for application of purified acetylene as an inhalation anaesthetic (3). The introduction of the highly combustible anaesthetic gas cyclopropane in 1933 urged anaesthetists to use fresh gas flows as low as possible to reduce pollution of the operating room and, thus, to minimize the risk of inadvertent explosions (4).
In 1954 halothane was introduced, a new volatile anaesthetic
characterized
by high anaesthetic potency yet narrow therapeutical width. To ensure
patients
safety, the use of this anaesthetic was bound to the knowledge of the
applied
vapour concentration. Its estimation, however, only was simple and easy
if a high fresh gas flow was used and the proportion of rebreathing was
kept rather low. The more, as the vaporizers, available at that time,
didn´t
work sufficiently reliably and precisely in the low flow range. Thus,
although
nearly all anaesthetic machines were already equipped with
sophisticated
rebreathing circle systems, paradoxically, it became clinical routine
to
use fresh gas flows as high as 4 to 6 l/min, completely excluding any
significant
rebreathing (4). In many countries this is still the routine way to
execute
inhalational anaesthesia (5). However, due to the development of modern
anaesthetic apparatus, the availability of comprehensive gas
monitoring,
an increasing environmental awareness, the introduction of new
advantageous
but expensive inhalational anaesthetics, and the world wide restriction
of economical ressources in medical care, since about fifteen years an
increasingly strong recollection towards low flow techniques can
be observed - and should be encouraged (6).
2. Low flow anaesthesia - the theory
Rebreathing systems can be used in different ways: If used with a fresh gas flow equal to the minute volume of the patient, the share of rebreathing will be neglectible. Nearly completely the expired air will vented out off the system as excess gas via the APL-valve. The patient gets nearly pure fresh gas. If a flow of 4.0 l/min is used, the share of rebreathing will increase to about 20%. The patient inhales a gas the composition of which still resembling that of the fresh gas. Only if the flow is reduced to at least 2.0 l/min or lower values, the share of rebreathing will reach 50% or more. Thus, only when low fresh gas flows are used the share of rebreathing will become significant, and judicious use is made frome the rebreathig technique (7).
According to the literature two different low flow techniques can be distinguished. The term Low Flow Anaesthesia was introduced by F. Foldes, inaugurating an anaesthetic technique performed with a fresh gas flow of 1.0 l/min (8). R. Virtue introduced the term Minimal Flow Anaesthesia by recommending the use of an even lower flow of 0.5 l/min (9). As emphasized beforhand, the lower the fresh gas flow the lower is the amount of gas vented out of the breathing system as waste and the higher is the proportion of rebreathing. The general term - low flow anaesthesia - should be restricted to defining an anaesthetic technique in which a semiclosed rebreathing system is used recirculating at least 50% of the exhaled air back to the patient after CO2 absorption. Using modern rebreathing systems this will be achieved only if the fresh gas flow is reduced to at least 2 l/min (10).
However, there is a limit for reducing the fresh gas flow: To prevent from gas volume deficiency at least that gas volume has to be delivered into the breathing system, which is definitely taken up by the patient (Fig.1).
During the course of anaesthesia oxygen is taken up constantly by the patient in the range of the basal metabolic needs. It can be calculated by applying a simplified version of Brody´s formula (11):
VO2 = 10 x BW [kg] 3/4 .
The uptake of nitrous oxide and the volatile anaesthetic, however, follows a power function. Nitrous oxide uptake of a normal body weight adult patient can be roughly estimated by applying Severinghaus´ formula (12)
VN2O = 1000 x t-1/2 ,
and the uptake of inhalational anaesthetics may be calculated by H. Lowe´s formula (13):
VAN = f x MAC x lB/G x Q x t-1/2 .
Thus, assuming a constant gas composition circulating within the breathing system, the total gas uptake, the sum of oxygen, nitrous oxide and inhalational anaesthetic uptake, is following a power function. Initially it is high and declines sharply during the first 30 minutes, but is comparatively low and decreases only delayed during the following time course of anaesthesia. This exponential characteristic of the gas uptake results from the fact that the partial pressure difference of anaesthetic gases between the alveolar space and the blood, being initially high, decreases continuously with increasing saturation of the blood and the tissues. If the anaesthetist, by frequent alterations of the settings at the gas controls, could succeed to approximate the total gas uptake, anaesthesia with closed rebreathing system would be realized. In clinical practice, however, continuous adaptations of the fresh gas flow according to the continuously changing individual gas uptake will be impossible. Whereas, applying very simple and safe standardized dosing schemes, low flow techniques like Minimal Flow and Low Flow Anaesthesia can be performed safely with already available anaesthetic equipment in routine clinical work (7,14).
3. Low flow anaesthesia - the practice
3.1 Induction
Premedication and induction of low flow anaesthesia is performed
according
to the usual induction scheme. Preoxigenation by applying pure oxygen
via
a face mask is followed by intravenous injection of a hypnotic. After
neuromuscular
relaxation and endotracheal intubation or insertion of a laryngeal mask
the patient is connected to the breathing system. In about 85% of all
cases
the gas tightness of the laryngeal mask will allow the fresh gas flow
to
be reduced to 0.5 l/min, even if controlled ventilation is performed
(7,15).
There are no procedure specific requirements for premedication and
induction.
3.2 Initial high flow phase
According to the guidelines given by Foldes or Virtue, during the
first
initial phase lasting 10 to 15 minutes a high fresh gas flow has to be
used. The author himself (7,16) recommends to set the oxygen flow to
1.4
l/min and the nitrous oxide flow to 3.0 l/min. This fresh gas
composition
guarantees in most of all patients an inspired oxygen concentration of
at least 30%, meeting the recommendations of Barton and Nunn (17,18).
The
following settings of the vaporizers are used routinely during the
initial
phase: enflurane 2.5 Vol%, isoflurane 1.5 Vol%, sevoflurane 2.5 Vol%,
and
desflurane 4.0 to 6.0%. If these settings are used over the first 10 to
15 minutes an expired concentration of about 0.7 to 0.8 times the MAC
of
the respective volatile agent will be gained. In addition to a nitrous
oxide MAC of about 0.6, corresponding to a nitrous oxide concentration
of 60%, this will result in a common MAC of 1.3 representing the AD95,
the anaesthetic gas concentration guaranteeing a sufficient anaesthetic
depth for 95% of all patients to tolerate the skin incision without any
movement. To initially use a high fresh gas flow is furthermore
indispensable
for sufficient denitrogenation and wash in of the aspired gas
composition
into the whole gas containing space. Last but not least, if the flow to
early would be reduced to too low values, inevitably gas volume
deficiency
would result compromizing adequate ventilation (7,16).
3.3 Flow reduction
If Low Flow Anaesthesia is to be performed, the fresh gas flow can be reduced to 1.0 l/min already after 10 minutes. Flow reduction will lead to a significant increase of rebreathing. The inspired gas, thus, contains a markedly increased proportion of the exhaled gas which already had passed the patient´s lung and contains less oxygen. The resulting decrease of oxygen content in the gas mixture has to be compensated by increasing the fresh gas oxygen content, which must be the higher the lower is the flow. Thus, to maintain a safe inspired oxygen concentration of about 30 % in Low Flow Anaesthesia, the fresh gas oxigen concentration has to be increased to 50%, but at least to 40%. With the fresh gas flow reduction, furthermore, the amount of anaesthetic vapour delivered into the system is markedly reduced. This has to be compensated by a corresponding significant increase of the agent´s concentration in the fresh gas. Only in this way the aspired anaesthetic concentration within the breathing system can be kept constant. In Low Flow Anaesthesia the fresh gas enflurane concentration is increased to 3.0 Vol%, isoflurane to 2.0 Vol% and sevoflurane to 3.0 Vol% (7,16,19). Due to its specific pharmacokinetic properties, only the fresh gas desflurane concentration can be maintained unchanged (20). Executing these standardized schemes, the expired anaesthetic concentrations will be maintained in the aspired range of 0.7 to 0.8 times the MAC.
If Minimal Flow Anaesthesia is to be performed, the initial high flow phase should last about 15 minutes. A sufficiently long initial high flow phase will prevent from accidental gas volume deficiency, which always will result whenever the gas loss via individual uptake and leakages is higher than the gas volume delivered into the system. To maintain a safe inspired oxygen concentration of at least 30%, oxygen fresh gas concentration has to be increased to 60%, but at least to 50%, when the flow is reduced to 0.5 l/min. Simultaneously the anaesthetic concentration of the fresh gas has to be increased: When enflurane is used to 3.5 Vol%, isoflurane to 2.5 Vol%, sevoflurane to 3.5 Vol%, and the Desflurane concentration standardizedly is raised by 1 Vol%. By applying this dosing scheme again an expired agent's concentration in the range of 0.7 to 0.8 times of the respective MAC will be maintained too (7,16,19,20).
3.3.1 Inspired oxygen and nitrous oxide concentration
After flow reduction from 4.4 to 0.5 l/min an initial increase of
the
FIO2 over the following period of 30 to 45 minutes can be observed. It
will be more pronounced in small or elderly patients with low oxygen
uptake
than in strong young and athletic patients. This initial increase is
followed
by a slow but continuous decline of the inspired oxygen concentration
to
lower values, on its part being the more pronounced the higher is the
oxygen
uptake of the patient (Fig.2). Whenever the lower alarm limit of the
oxygen
monitoring is reached, which carefully must be adjusted to 30%, the
oxigen
flow has to be increased by 10% of the total fresh gas flow, whereas
the
nitrous oxide flow must be decreased by the same amount. Thus, in
Minimal
Flow Anaesthesia, the oxygen flow has to be increased by 50
ml/min
and the nitrous oxide flow to be reduced by 50 ml/min likewise. After
these
adjustments, again, first a slow increase of the FIO2 will be
followed
by its slow but continuous decrease. Whenever the lower alarm threshold
is reached anew, the oxygen flow again has to be increased by 10% of
the
total fresh gas flow and the nitrous oxide flow to be reduced by the
same
amount. In Low and Minimal Flow Anaesthesia the oxygen concentration
within
the breathing system is changing slowly but continuously during the
course
of the anaesthetic procedure (7,16).
3.3.2 Concentration of inhalational anaesthetics
If, with reduction of the flow, the fresh gas concentration of the
volatile
anaesthetic is increased according to the given standardized schemes, a
slight decrease of the inspired and expired anaesthetic concentration
can
be observed. The vaporizers of all modern anesthetic apparatus are
switched
into the fresh gas line (VOC). Thus, the reduction of the flow results
in a corresponding significant decrease of the amount of anaesthetic
vapour
delivered into the system. In all the different anaesthesia machines,
likewise,
the breathing system, the ventilator, the connecting hoses and the
patient
hose assembly contain a gas volume of about 5 to 6 liter. This, in
addition
to the gas volume of about 2.5 liter contained in the lung of an adult
patient, is the distribution space for the anaesthetic vapour delivered
into the system (7). A fresh gas flow as low as 500 ml/min assumed, a
change
of the vaporizer's setting from zero to 5 Vol% only will raise the
amount
of vapour from 0 to 25 ml/min, a considerably small volume if compared
with the distribution space. Thus, in low flow anaesthesia there is a
marked
difference between the anaesthetic´s fresh gas
concentration
and its concentration within the breathing system, and this difference
is the higher the lower is the fresh gas flow, but the lower the less
soluble
is the anaesthetic agent (19-21) (Fig.3). If the concentration of the
volatile
anaesthetic shall be changed, the vaporizer has to be adjusted to a
concentration
considerably exceeding the aspired nominal value.
3.3.3 Time constant
The time constant is a measure for the time it takes, that alterations of the fresh gas composition will lead to corresponding alterations of the gas composition within the breathing system. According to a formula given by Conway (22) the time constant (T) can be calculated by the division of the system's volume (VS) by the difference between the amount of anaesthetic agent delivered into the system with the fresh gas (VD) and the individual gas uptake (VU):
T = VS / (VD - VU) .
A given volume of the system and a given individual gas uptake
assumed,
the time constant is reversely proportional to the fresh gas flow. The
marked increase of the time constant has to be taken into account when
switching from high to low fresh gas flows (Fig. 4). Whenever the gas
composition
within the breathing system needs to be changed rapidly, the fresh gas
flow has to be increased for adequately accelerating the wash in of the
newly aspired gas composition. If low flow anaesthesia is performed
with
the newer volatiles, characterized by low anaesthetic potency and
solubility
like sevoflurane and desflurane, the time constants will be
significantly
shorter as VD can be raised considerably and VU is extremely low
(19-21,23).
3.4 Recovery phase
According to the long time constant, the vaporizer can be closed
about
15 to 20 minutes before the definite end of the surgical procedure. If
the low flow is maintained, the decrease of the anaesthetic's
concentration
is delayed and slow. During that time recovery of spontaneous breathing
can be induced by using the SIMV ventilation mode or by manual
assistance
of the ventilation. Not until about five minutes before extubation the
anaesthetic gases are washed out by switching to high flow of pure
oxigen.
Care is taken of the recovering patient in the usual manner (7,16).
3.4.5 Characteristics of low flow anaesthesia
If commercially available anaesthetic machines are used, with Low
and
Minimal Flow Anaesthesia the maximum of flow reduction is reached which
can be gained in routine clincal practice. Both techniques are extreme
variants of semiclosed use of rebreathing systems, as still a small
amount
of excess gas is used.
The performance of Low and Minimal Flow Anaesthesia becomes very simple
if standardized schemes are used to controll the fresh gas flow and its
composition. This schemes require only rare adjustments at the gas flow
controls and vaporizers. The anaesthetist, however, must accept that
the
gas concentrations within the breathing system not will remain constant
at the aspired values, but rather will change slowly but continuously
during
the course of anaesthesia.
Last but not least, standardized schemes for the performance of low flow anaesthesia only can be guidelines. The fresh gas flow and its composition always must be adapted to the individual patient's reaction and the current requirements of the surgical procedure.
4. Technical preconditions for safe performance of low flow
anaesthesia
4.1. Monitoring and alarm thresholds
Due to its specific characteristics following monitoring is
essential
for safe performance of low flow anaesthesia (6,7,16): As the
difference
between the gas concentrations in the breathing system and the fresh
gas
increases with the extend of flow reduction, the anaesthetic gas
composition
can't be reliably assessed from the composition of the fresh gas. Thus,
continuous monitoring of the inspired oxigen concentration is
absolutely
indispensable. The same applies for the concentration of volatile
anaesthetics,
if a fresh gas flow lower than 1 l/min is used. The lower threshold for
the inspired oxigen concentration should be set to 30%, and the upper
alarm
limit of inspired anaesthetic concentration to 2.0-3.0 Vol% for
Halothane,
Enflurane and Isoflurane likewise, to 5.0 Vol% for Sevoflurane and 8.0
Vol% for Desflurane. Always the fresh gas volume must be as large as to
sufficiently compensate the gas loss via individual uptake and
leakages.
Otherwise gas volume deficiency will occure, inevitably leading to an
alteration
of the ventilation. Continuous monitoring of the air way pressure or,
alternatively,
the minute volume therefore also is indispensable. The disconnect alarm
should be set to a value 5 mbar lower than the peak pressure, the lower
alarm limit of the minute volume monitoring to 0.5 l/min lower than the
aspired minute volume. If low flow techniques are performed
consistently
the soda lime consumption will increase fourfold. By continuous
monitoring
of the inspired carbon dioxide concentration soda lime exhaustion
reliably
can be detected. If this monitoring is not available jumbo or double
absorber
canisters should be used and the soda lime changed after each day of
work
(7,24).
4.2. Anaesthetic apparatus
The technical features of the anaesthetic apparatus have to comply with following requirements (7,10):
The flow control system must feature needle valves and flow meter tubes calibrated and reliably working even in the low flow range, the vaporizers must be fresh gas flow compensated. The rebreathing system has to be sufficiently gas tight: The leackage must not exceed 100 ml/min at a pressure of 20 mbar to meet the requirements for Minimal Flow Anaesthesia. The performance of low flow techniques is significantly facilitated by the availability of an anaesthetic gas reservoir, by which small accidental gas volume deficiencies can be balanced. Such gas reservoir can be, alternatively, the endinspiratory volume contained in the bag of a bag-in-bottle ventilator, or the bellows of a ventilator with standing or hanging „floating“ bellows, or the manual ventilation bag in machines equipped with a fresh gas decoupling valve. If an anaesthetic apparatus is used, featuring continuous flow of the fresh gas into the breathing system, furthermore, it has to be considered that each alteration of the fresh gas flow will lead to a corresponding alteration of the tidal volume. Only if an anaesthetic ventilator is used, featuring fresh gas flow compensation, the preset tidal volume will be delivered independently from the fresh gas flow range. Fresh gas flow compensation can be realized, alternatively, by discontinuous delivery of the fresh gas into the breathing system, or by automatic electronical control of the ventilator´s performance corresponding to the fresh gas flow rate.
Using modern anaesthetic equipment even in pediatric anaesthesia low
flow techniques are advantageous and can be executed safely (7,25,26).
5. Advantages of low flow anaesthesia
The advantages of low flow anaesthesia are obvious and indisputable and were already comprehensively listed in Waters´ paper (2): the reduction of anaesthetic gas and vapour consumption, the decrease of atmospheric pollution with inhalation anaesthetics, the improvement of anaesthetic gas climate, and the significant reduction of costs.
Comparing two hours of continuous high flow (4.5 l/min) with minimal flow (0.5 l/min) isoflurane anaesthesia, the consumption of oxygen is reduced by 115 l, that of nitrous oxide by 300 l, and that of isoflurane vapour by 5.6 l (7). If high flow techniques, using fresh gas flows of about 4.5 l/min, would be replaced consistently by low flow anaesthesia, in Germany and the UK the resulting reduction in gas and anaesthetic vapour consumption were projected to be about 350 million liters of oxygen, 1000 million liters of nitrous oxide, 33500 liters of fluid isoflurane and 46250 liters of fluid enflurane (10). The conclusion is very simple and obvious: The lower the flow the less the gas consumption (27,28).
Anaesthetists also have to deal with increasingly stringent official regulations on the maximum acceptable workplace concentrations of anaesthetic gases (29). Careful maintenance of the anaesthetic apparatus and scrupulous attention on leaks from breathing systems provided, even the extremely low anaesthetic gas concentrations stipulated by the US National Institute of Occupational Safety and Health can be achieved easily only by the use of low flow techniques (30,31). Most operating theatres, however, are equipped with central gas-scavenging systems, and it is possible to stay within the defined limits even if high fresh gas flows are used. Nevertheless, high flow anaesthesia will inevitably result in pollution of the atmosphere beyond the operating theatre. Both, nitrous oxide and the volatile anaesthetics contribute to the destruction of the ozone layer and to the greenhouse effect. The ozone destructive potential of the volatile anaesthetics halothane, enflurane and isoflurane, which are partially halogenated chlorofluorocarbons (CFCs), is assumed to be only 0.1-1 % of all fully substituted CFCs. Furthermore, the proportion of nitrous oxide, emitted from hospitals, is only about 1 % of the total amount of nitrous oxide polluting the atmosphere. Most is derived from bacterial metabolism in fertilised soil. Nevertheless, even if the emission of anaesthetic gases is a comparatively small fraction of the total polluting gases, anaesthetists are morally obliged to minimize pollution in an age of increasing environmental awareness, and it is their duty to use all technical facilities available to achieve this (32-36). Desflurane and sevoflurane, which are halogenated only with fluorine, are assumed to have nearly no ozone depleting potential but may contribute considerably to the greenhouse effect.
Appropriate humidification and warming of anaesthetic gases have
significant
impact on the function and the integrity of the ciliated epithelium of
the respiratory tract. During anaesthesia, the absolute humidity of the
inspired gas should range preferably between 17-30 mgH2O/l, and its
temperature
between 28-32 °C. After an initial period of 30-45 minutes in
an actively heated compact breathing system, the values referred to
above
can be achieved solely by the use of a low flow technique (37-41).
Cost savings directly result from the decrease of gas and anaesthetic
consumption. They are related to the duration of the anaesthetic
procedure,
the price of the respective anaesthetic agent and the extend of flow
reduction
(42-44). Comparing high flow (4.5 l/min) with a minimal flow (0.5
l/min)
technique lasting 2 hours, assuming the inspired anaesthetic
concentration
at MAC, savings of about US$ 15 can be gained if enflurane, about US$
21
if isoflurane and about US$ 47 if desflurane is used (16). By
comparison,
the additional cost of about US$ 0.60 resulting from increased
consumption
of soda lime during 2 hours is negligible (7). Relating to the
beforehand
mentioned projection, the annual financial savings resulting from
reduced
gas and anaesthetic consumption in Germany and UK are assumed to total
more than US$ 65,36 million if low flow anaesthetic techniques would be
performed consistently (10). It seems to be realistic to assume a cost
saving in the range of 50 - 75% if low flow techniques would be used
consistently
in clinical routine practice (7,45,46). Even the more costly
anaesthetics
like desflurane could be used without significant increase in costs
(47).
5.1 Efficiency of inhalation anaesthetic techniques
One of the most striking aguments in favour for low flow anaesthesia
is the marked increase in efficiency of inhalational anaesthesia (48).
The efficiency (Eff) can be calculated by dividing the amount of agent
taken up by the patient (VU) by the amount of agent delivered into the
breathing system (VD):
Considering this algorithm it becomes obvious that an inhalation
anaesthetic
technique is the less efficient, the lower is the individual uptake and
the higher is the amount of agent which is delivered into the breathing
system. As the amount of agent delivered into the system directly is
bound
to the fresh gas flow, the anaesthetic technique will be the less
efficient
the higher is the fresh gas flow. And this, especially, holds for
anaesthetic
agents featuring low solubility and anaesthetic potency (Fig. 5). If
desflurane,
for instance, is used with a flow of 4.5 l/min at an inspired
concentration
of 6.0 % over a period of 2 hours, the overall efficiency will
decline
to 0.07: Only 7 % of the total amount of agent delivered into the
system
is really needed and taken up by the patient, whereas 93 % are wasted
with
the excess gas escaping from the breathing system (7). Only if this
agent
is applied with low fresh gas flow rates, the efficiency can be
increased
to an acceptable range of about 30%. The use of anaesthetic agents
featuring
likeweise low solubility and anaesthetic potency, like Sevoflurane and
Desflurane, for economic and ecologic reasons only can be justified if
judicious use is made of rebreathing techniques (47,49,50).
6. Trace gas accumulation
A matter of concern remains the accumulation of trace gases resulting from the diminution of the wash out of foreign gases, which is the less the lower is the fresh gas flow. Foreign gases may decrease the concentration of nitrous oxide and oxygen. That may, for instance, be the case if nitrogen accumulates due to insufficient denitrogenation, or the argon concentration may rise due to the use of an oxygen concentrator (51,52). Methan, physiologically exhaled by the patient, in high concentrations may compromise the measurement and monitoring of halothane concentration (53). Accumulation of acetone may prolong the emergence from anaesthesia and provoke nausea or vomiting. However, only in the very rare cases of severely ketoacidotic patients this really may become clinically relevant (54). Accumulation of these trace gases even in prolonged low flow anaesthesia, up to now, have not been shown to be of real clinical importance (55).
All inhalational anaesthetics react with carbon dioxide absorbents
by
absorption and degradation, most eagerly if the absorbent is desiccated
(56). The new volatiles desflurane and sevoflurane are more liable to
react
with the alkaline absorbents than the older anaesthetics
halothane,
enflurane or isoflurane. Desflurane more than enflurane and isoflurane
react with absolutely dry carbon dioxide absorbents by generating
carbon
monoxide. Only partial wetting markedly reduces this chemical reaction,
and if soda lime contains only 4.8% and Baralyme only 9.5% water,
carbon
monoxide generation is suppressed completely (57). It had been
concluded,
consistently not to use fresh gas flows lower than 5 l/min to safely
avoid
accidental carbon monoxide intoxication resulting from trace gas
accumulation
(58). This conclusion, however, strongly must be rejected, as just high
fresh gas flows are liable to dry out the absorbents. Contrarily, low
flow
anaesthesia, preserving the moisture content of the absorbents, just
can
be taken as a measure preventing from carbon monoxide generation (59).
Sevoflurane more eagerly than halothane reacts with dry absorbents too
(60,61). Both agents, however, also react with normally wet carbon
dioxide
absorbents by generating haloalkenes: Halothane by forming BCDFE =
1,bromo-1,chloro-2,2,difluoro-ethylene
(62) and sevoflurane by forming compound A =
fluoromethyl-2,2,difluoro-1,trifluoromethyl-vinylether.
Compound A concentration was found to increase with the extend of flow
reduction, with the absorbent´s temperature and the agent´s
concentration (63,64). Some authors regard a compound A load of 150 to
240 ppmh as potentially nephrotoxic in humans (65-67) and emphasize not
to use this anaesthetic with flows lower than 2.0 l/min. Contrarily,
some
authors estimate Low Flow Anaesthesia with sevoflurane to be safe,
arguing
that mean peak concentrations in different studies did not exceed 25
ppm
and no signs of renal impairment were observed in any patient (49,
68,69).
Mazze recently published the results of an investigation on
nephrotoxicity
of compound A in primates demonstrating that only at a load of at least
800 ppmh nephrotoxic effects occured (70). Accepting this threshold for
nephrotoxic load with compound A absolutely no flow restriction
would
be justified. Even long lasting Minimal Flow Anaesthesia with
sevoflurane
could be performed safely, although compound A peak concentrations were
found to reach 50 to 60 ppm with this technique (71). Unlike in the
United
States, sevoflurane was approved for clinical use without any fresh gas
flow restriction in all countries of the European Common Market.
Nevertheless,
whenever in clinical practice there might be assumed the possibility of
accumulation of potentially harmful trace gases, for safety reasons, a
low flow technique using a flow of at least 1 l/min should be
performed,
guaranteeing a sufficient continuous wash out effect (7).
7. Literature
1. Snow J. On Narcotism by the Inhalation of Vapours. Part XV. The effects of Chloroform and Ether Prolonged by Causing the Exhaled Vapour to be reinspired. London Medical Gazette 1850; 11: 749-754
2. Waters RM. Clinical Scope and Utility of Carbon Dioxid Filtration in Inhalation Anaesthesia. Anesth Analg 1924; 3: 20-2
3. Baum JA. Who introduced the Rebreathing System into Clinical Practice? In Schulte am Esch J., M. Goerig, eds. Proceedings of the Fourth International Symposium on the History of Anaesthesia. Lübeck: Dräger 1998: 441-450
4. Onishchuk, JL. The early History of Low-Flow Anaesthesia. In Fink, B.R., L.E. Morris, C.R. Stephen, eds. The History of Anesthesia. Third International Symposium, Proceedings. Park Ridge, Illinois: Wood Library-Museum of Anesthesiology, 1992: 308-313
5. Cravero J, Suida E, Manzi DJ, Rice LJ. Survey of low flow anesthesia in the United States. Anesthesiology 1996; 85: A995
6. Baxter A. Low and minimal flow inhalation anaesthesia. Can J Anaesth 1997; 44: 643-653
7. Baum JA. Low Flow Anaesthesia. The Theory and Practice of Low
Flow,
Minimal Flow and Closed System Anaesthesia. Oxford: Butterworth
Heinemann,
1996
8. Foldes FF, Ceravolo AJ, Carpenter SL. The administration of nitrous oxide - oxigen anesthesia in closed systems. Ann Surg 1952; 136: 978-981
9. Virtue RW. Minimal flow nitrous oxide anesthesia. Anesthesiology 1974; 40: 196-198
10. Baum JA, Aitkenhead AR. Low-flow anaesthesia. Anaesthesia 1995; 50 (Suppl.): 37-44
11. Kleiber M. Body size and metabolic rate. Physiol Rev 1945; 27: 511-539
12. Severinghaus J W. The rate of uptake of nitrous oxide in man. J Clin Invest 1954; 33: 1183-1189
13. Lowe H J, Ernst E A. The Quantitative Practice of Anesthesia. Williams & Wilkins, Baltimore 1981
14. Baker AB. Back to the basics - a simplified non-mathematical approach to low flow techniques in anaesthesia. Anaesth Intensive Care 1994; 22: 394-395
15. Fröhlich D, Schwall B, Funk W, Hobbhahn J. Laryngeal mask airway and uncuffed tracheal tubes are equally effective for low flow or closed system anaesthesia in children. Br J Anaesth 1997; 79: 289-292
16. Baum JA. Low Flow Anaesthesia with Dräger Machines. Questions and Answers. 3rd ed. Lübeck, Dräger Medizintechnik, 1998
17. Barton F, Nunn JF. Totally closed circuit nitrous oxide/oxygen anaesthesia. Br J Anaesth 1975; 47: 350-357
18. Don H. Hypoxemia and hypercapnia during and after anesthesia. In: Orkin F K, Cooperman L H, eds. Complications in Anesthesiology. Lippincott, Philadelphia 1983, pp. 183- 207
19. Baum J, Stanke HG. Low flow and minimal flow anaesthesia with sevoflurane (German). Anaesthesist 1998; 47 (Supplement 1): S70-S76
20. Baum J, Berghoff M, Stanke HG, Petermeyer M, Kalff G. Low-flow anaesthesia with desflurane (German). Anaesthesist 1997; 46: 287-293
21. Hargasser S, Hipp R, Breinbauer B, Mielke L, Entholzner E, Rust M. A lower solubility recommends the use of desflurane more than isoflurane, halothane, and enflurane under low-flow conditions. J Clin Anesth 1995; 7: 1-5
22. Conway CM. Closed and low flow systems. Theoretical considerations. Acta Anaesth Belg 1984; 34: 257-263
23. Avramov MN, Griffin JD, White PF. The effect of fresh gas flow and anaesthetic technique on the ability to control hemodynamic responses during surgery. Anesth Analg 1998; 87: 666-670
24. Morris LE. Closed carbon dioxide filtration revisited. Anaesth Intensive Care 1994; 22: 345-358
25. Peters JWB, Bezstarosti J, van Eden, Erdman W, Meursing AEE. Safety and efficacy of semi-closed circle ventilation in small infants. Pediatric Anaesthesia 1998; 8: 299-304
26. Igarashi M, Watanabe H, Iwasaki H, Namiki A. Clinical evaluation of low-flow sevoflurane anaesthesia for pediatric patients. Acta Anaesth Scand 1999; 43: 19-23
27. Feiss P, Demontoux MH, Colin D. Anesthetic gas and vapour saving with minimal flow anesthesia. Acta Anesth Belg 1990; 41: 249-251
28. Pedersen FM, Nielsen J, Ibsen M, Guldager H. Low-flow isoflurane-nitrous oxide anaesthesia offers substantial economic advantages over high-flow and medium flow isoflurane-nitrous oxide anaesthesia. Act Anaesth Scand 1993; 37: 509-512
29. Spence AA. Environmental pollution by inhalation anaesthetics. Br J Anaesth 1987; 59: 96-103
30. Virtue RW. Low flow anesthesia: advantages in its clinical application, cost and ecology. In: Aldrete JA, Lowe HJ, Virtue RW, eds. Low Flow and Closed System Anesthesia. Grune & Stratton, New York 1979, pp. 103-108
31. Imberti R, Preseglio I, Imbriani M, Ghittori S, Cimino F, Mapelli A. Low flow anaesthesia reduces occupational exposure to inhalation anaesthetics. Acta Anaesth Scand 1995; 39: 586-591
32. Logan M, Farmer JG. Anaesthesia and the ozone layer: Br J Anaesth 1989; 53: 645-646
33. Noerreslet J, Frieberg S, Nielsen TM, Römer U. Halothane anaesthetic and the ozone layer. Lancet 1989; 719
34. Pierce JMT, Linter SPK. Anaesthetic agents and the ozone layer. Lancet 1989; 1011-1012
35. Sherman SJ, Cullen BF. Nitrous oxide and the greenhouse effect.
Anesthesiology 1988; 68: 816-817
36. Solomon S, Albritton D. Time-dependent ozone depletion potentials
for short and long-term forecasts. Nature 1992; 357: 33-37
37. Bengtson JP, Sonander H, Stenqvist O. Preservation of humidity and heat of respiratory gases during anaesthesia - a laboratory investigation. Acta Anaesthesiol Scand 1987; 31: 127-131
38. Bengtson JP, Bengtson A, Stenqvist O. The Circle System as a Humidifier. Br J Anaesth 1989; 63: 453-457
39. Branson RD, Campbell RS, Davis K, Porembka DT. Anesthesia circuits, humidity output, and mucociliary structure and function. Anesth Intens Care 1998; 26: 178-183
40. Chalon J, Ali M, Turndorf H, Fischgrund GK. Humidification of anesthetic gases. Charles C. Thomas, Sprigfield 1981
41. Kleemann PP. Humidity of Anesthetic Gases with Respect to Low Flow Anaesthesia. Anaesth Intens Care 1994; 22: 396-408
42. Bengtson JP, Sonander H, Stenqvist O. Comparison of costs of different anaesthetic techniques. Acta Anaesth Scand 1988; 32: 33-35
43. Christensen KN, Thomsen A, Jorgensen S, Fabricius J. Analysis of Costs of Anaesthetic Breathing Systems. Br J Anaesth 1987; 59: 389-390
44. Loke J, Shearer WAJ. Cost of anaesthesia. Can J Anaesth. 1993; 40: 472-474
45. Cotter SM, Petros AJ, Doré CJ, Berber ND, White DC. Low-flow anaesthesia. Anaesthesia 1991; 46: 1009-1012
46. McKenzie AJ. Reinforcing a „low flow“ anaesthesia policy with feedback can produce a sustained reduction in isoflurane consumption. Anaesth Intensive Care 1998; 26: 371-376
47. Eger EI. Economic analysis and pharmaceutical policy: a consideration of the economics of the use of desflurane. Anaesthesia 1995; 50 (Supplement): 45-48
48. Ernst EA, Spain JA. Closed-circuit and high-flow systems:
Examining
alternatives. In: Brown BR, ed. Future Anesthesia Delivery Systems.
Contemporary
Anesthesia Practice, Vol. 8. F. A. Davies, Philadelphia 1984, pp. 11-38
49. Mazze RI, Jamison RL. Low-flow (1 l/min) sevoflurane - Is it safe?.
Anesthesiology 1997; 86: 1225-1227
50. Smith I, Nathanson M, White PF. Sevoflurane - a long-awaited volatile anaesthetic. Brit J Anaesth 1996; 76: 435-445
51. Morita S, Latta W, Hambro K, Snider M. Accumulation of methane, acetone and nitrogen in the inspired gas during closed-circuit anesthesia. Anesth Analg 1985; 64: 343-347
52. Parker CJR, Snowdon SL. Predicted and measured oxygen concentrations in the circle system using low fresh gas flows with oxygen supplied by an oxygen concentrator. Br. J Anaesth 1988; 61: 397-402
53. Rolly G, Versichelen LF, Mortier E. Methane Accumulation During Closed-Circuit Anesthesia. Anesth Analg 1994; 79: 545-547
54. Strauß JM, Hausdörfer J. Accumulation of Acetone in Blood During Long-Term Anaesthesia with Closed System. Br J Anaesth 1993; 70: 363-364
55. Baumgarten RK, Reynolds WJ. Much Ado About Nothing: Trace Gaseous Metabolites in the Closed Circuit. Anesth Analg 1985; 64: 1029-1030
56. Strum DP, Eger II EI. The degradation, absorption and solubility
of volatile anesthetics in soda lime depends on the water
content.
Anesth Analg 1994; 78: 340-348
57. Fang ZX, Eger II EI, Laster MJ, Chortkoff BS, Kandel L, Ionescu
P. Carbon monoxide production from degradation of desflurane, enflurane
isoflurane, halothane and sevoflurane by soda lime and baralyme. Anesth
Anlag 1995; 80: 1187-1193
58. Moon RE. Carbon monoxide gas may be linked to CO2 absorbent. Anesth Patient Safety Found Newslett 1991; 6: 8
59. Baum J, Sachs G, v. d. Driesch C, Stanke HG. Carbon monoxide generation in carbon dioxide absorbents. Anesth Analg 1995; 81: 144-146
60. Baum J, Sitte Th, Strauß JM, Forst H, Zimmermann H, Kugler B. Absorption and degradation of sevoflurane in dry soda lime (German). Anästh Intensivmed 1998; 39: 11-16
61. Funk W, Gruber M, Wild K, Hobbhahn J. Dry soda lime markedly degrades sevoflurane during simulated inhalation induction. Br J Anaesth 1999; 82: 193-198
62. Sharp HJ, Trudell JR, Cohen EN. Volatile metabolites and decomposition products of halothane in man. Anesthesiology 1979; 50: 2-8
63. Fang ZX, Eger II EI. Factors affecting the concentration of compound A resulting from degradation of sevoflurane by soda lime and Baralyme Ò in a standard anesthetic circuit. Anesth Analg 1995; 81: 564-568
64. Fang ZX, Kandel L, Laster MJ, Ionescu P, Eger II EI. Factors affecting production of compound A from the interaction of sevoflurane with Baralyme Ò and soda lime. Anesth Analg 1996; 82: 775-781
65. Eger EI, Gong D, Koblin DD, Bowland T, Ionescu P, Laster MJ, Weiskopf RB. Dose-related biochemical markers of renal injury after sevoflurane versus desflurane anesthesia in volunteers. Anesth Analg 1997; 85: 1154-1163
66. Eger EI, Gong D, Koblin DD, Bowland T, Ionescu P, Laster MJ, Weiskopf RB. The effect of anesthetic duration on kinetic and recovery characteristics of desflurane versus sevoflurane, and on the kinetic characteristics of compound A in volunteers. Anesth Analg 1998; 86: 414-421
67. Goldberg ME, Cantillo J, Gratz I, Deal E, Vekeman D, McDougall R, Afshar M, Zafeiridis A, Larijani G. Dose of compound A, not sevoflurane, determines changes in the biochemical markers of renal injury in healthy volunteers. Anesth Analg 1999; 88: 437-445
68. Bito H, Ikeuchi Y, Ikeda K: Effects of low-flow sevoflurane anesthesia on renal function. comparison with high-flow sevoflurane and low-flow isoflurane anesthesia. Anesthesiology 86 (1997) 1231-1237
69. Kharash ED, Frink EJ, Zager R, Bowdle TA, Artru A, Nogami WM: Assessment of low-flow sevoflurane and isoflurane effects on renal function using sensitive markers of tubular toxicity. Anesthesiology 86 (1997) 1238-1253
70. Mazze RI, Friedman M, Delgado-Herrera L, Galvez ST, Mayer DB. Renal toxicity of compound A plus sevoflurane compared with isoflurane in non-human primates. Anesthesiology 1998; 89: A490
71. Reinhardt C, Gronau E, Wüsten R, Goeters C, Vrana S, Baum
J,
van Aken H. Compound A in Minimal Flow Sevoflurane. Anesthesiology
1998,
89 (No 3A): A142